Current output exchange Phase current output (effective value): 0-40A/per phase Triple current output (effective value): 0-120A Long term allowable working value of phase current (effective value): 10A/per phase Maximum output capacity of phase current: 450VA Maximum capacity at 120A triple current: 1200VA Allowable working time at 120A triple current: 8s Frequency range (fundamental): 20-1000Hz Harmonic frequency: 1-9 times
direct current Phase current output: 0- ± 20A/per phase Maximum output load voltage: 20V
Voltage output exchange Phase voltage output (effective value): 0-130V Line voltage output (effective value): 0-260V Phase voltage/line voltage output power: 80VA/160VA Frequency range (fundamental): 20-1000Hz Harmonic frequency: 1-9 times
direct current Phase voltage output amplitude: 0- ± 180V Line voltage output amplitude: 0- ± 360V Phase voltage/line voltage output power: 80VA/160VA
Switching input Empty contact: 1-20mA, 24V Potential contact input: "0": 0-+6V; '1':+11v -+250V
Switching output Empty contact point: DC: 220V/0.2A; AC: 220V/0.5A
Output and measurement accuracy
Accuracy within the rated range
current AC phase current:<0.2% ± 0.1% FS (FS is the full scale value) AC high current terminal output:<0.2% ± 0.1% FS DC current:<0.5% ± 0.1% FS
voltage AC voltage:<0.2% ± 0.1% FS (FS is the full scale value) DC voltage:<0.5% ± 0.1% FS
other Frequency accuracy (at 50Hz):<0.001 Hz ± 0.1% FS (FS is the full scale value) Phase accuracy:<0.5 ° Harmonic distortion:<0.5% Time measurement accuracy:<0.1ms
Time measurement range: 0.1ms - 9999s
Permissible range of power supply voltage: AC 220V ± 10%, 50/60Hz
Allowable range of ambient temperature: -10 -+50
Size: 450mm×235mm×360mm
Weight: 15kg